Page 11
Air Techniques, Inc.
View A . Verify Vertical Leveling
View B . Verify Horizontal Leveling
Site Survey
Make sure that the wall for mounting is strong enough for the installation and meet the support
load requirements of
paragraph 4 .6 .
Make sure to review and understand all Technical data provided by paragraph 3 .1 above .
Make sure the mounting wall is level in both vertical and horizontal directions using a level
indicator as shown below . If the wall is not level, adjust the leveling of the unit to be installed
by inserting shims as required between wall plate and base unit plate .
Make sure there shall be no electrical wiring conduits running around the area to be drilled
for the installation bolts .
Make sure that control unit to be mounted on the wall at a height of 45 .25 inches (1150mm) for
single stud mounting from the floor (above finished floor) . However this can change based on
site conditions without affecting the functionality of the system .
Make sure the location allows sufficient space for movement of the arms in the extended condition .
Make sure that the location allows the Provecta HD to be used with ease for all possible imaging
procedures on the patient with respect to the patient chair location .
Site Environment Requirements
Provecta HD is designed for indoor usage .
It should not be subjected to direct sunlight for any extended duration .
Mount the Provecta HD unit away from sources of liquid ingress .
If the unit is stored below 50°F (10°C) , time must be allowed for the unit to reach room tem-
perature before connecting it to the mains voltage .
Electrical Outlets & Requirements
The mains outlet should have a ground connection . Grounding of the electrical system must be
checked before connecting Provecta HD .
Additional wiring required for the site must done by a qualified electrician . All wiring should
conform to requirements provided by the User manual and in accordance with local codes .
The mains outlet should be capable of supplying 16A (110V) of current . It shall have fuse
protection or provided with a circuit breaker of 16A (110V) .
It is recommended to have an ELCB (Earth Leakage Circuit Breaker) for protection against
earth leakage .
4.6 Site Preparation