14
MODE
DEV TEMP
DISPLAY
READY
LIGHT
NORM
ARROW
ENDO
ARROW
PROC
LIGHT
SHUTTER HEATER POWER
DRYER
&
RECIRC
REPLEN
PUMP
Turn ON
> -15° F
Developer
Temperature
Heatup
OFF
ON
OFF
OFF
Closed
100%
OFF
ON
(90 sec)
Turn ON
< 77° F
Developer
Temperature
Heatup
OFF
ON
OFF
OFF
Closed
100%
OFF
ON
(90 sec)
Turn ON
< 77° F
Developer
Temperature
Heatup
OFF
or
ON
ON
OFF
OFF
Closed
100% (1st min)
<100% (to OFF)
OFF
ON
(90 sec)
Normal
Temperature
Rise
77 to 81° F
Developer
Temperature
Heatup
OFF
ON
OFF
OFF
Closed
90 to 20%
OFF
OFF
Fast
Temperature
Rise
77 to 81° F
Developer
Temperature
Heatup
OFF
ON
OFF
OFF
Closed
45 to 10%
OFF
OFF
At Set Point
82° F
82° F
ON
ON
OFF
OFF
Closed
OFF
OFF
OFF
In Control Loop
(Guard Band)
SEE NOTE 1
81 to 84° F
80 to 83° F
82° F
(ADC=118
(ADC=118)
SEE NOTE 2
ON
ON
OFF
OFF
Closed
OFF to 20% &
100% for 1 min
every ADC change
if temp drops.
SEE NOTE 3
OFF
OFF
Developer Hot
> 83° F
Developer
Temperature
OFF
ON
OFF
OFF
Closed
OFF
OFF
OFF
Processing
NORMAL
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
Flashing
Open
(+5 sec)
Off to 20%
ON
ON as long
as shutter
is open
SEE NOTE 4
Processing
NORMAL
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
ON
Closed
Off to 20%
ON
ON as long
as shutter
is open
SEE NOTE 4
Processing
ENDO
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
Flashing
Open
(+5 sec)
Off to 20%
ON
ON as long
as shutter
is open
SEE NOTE 4
Processing
ENDO
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
ON
Closed
Off to 20%
ON
ON as long
as shutter
is open
SEE NOTE 4
Processing
CLEANING
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
Flashing
(short/long)
Closed
Off to 20%
ON
OFF
Processing
CLEANING
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
Flashing
(short/long)
Open
Off to 20%
ON
OFF
Processing
CLEANING
(shutter open fail)
80 to 84° F
84° F
(Set Point)
ON
ON
OFF
Flashing
(short/long)
Open, but
no signal =
Closed
Off to 20%
ON
OFF
FUNCTION CHARTS
Notes:
1.
Control loop 80 to 84°F during heat up and from above 84°F upon cool down.
2.
Set point temperature range is between 79 and 86°F (ADC = Analog to Digital Count).
3.
100% HEATER POWER for 1 minute when processor power is turn ON each time the ADC changes
(increases) as temperature drops and every 4 minutes if temperature is below set point and ADC does
not change.
4.
See page 17 for pump on times.
FUNCTIONAL MODES
Содержание A/T2000 XR
Страница 1: ...AUTOMATIC FILM PROCESSOR SERVICE MANUAL...
Страница 20: ...20 SYSTEM SCHEMATIC...
Страница 40: ...40 BASE ASSEMBLY TOP VIEW OVERHEAD VIEW DETAIL A BASE ASSEMBLY 45200...
Страница 42: ...42 BASE ASSEMBLY BOTTOM VIEW SINGLE FUSE CONFIGURATION 5A...
Страница 49: ...49 ROLLER TRANSPORTS WASH DRY TRANSPORT P N5510 DEVELOPER TRANSPORT P N45500 FIXER TRANSPORT P N 45505...