13
FUNCTION CHARTS
AT2000 XR FUNCTION & OPERATION CHARTS
The AT2000 XR Film Processor's function and status conditions, as designed, are separated
into six categories listed below. The charts associated will provide insite to required processor
operation and expedite any trouble-shooting, if applicable. Refer to the appropriate Function
Chart, as required, to quickly identify and resolve the issue at hand.
q
FILM SHUTTER FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
q
FUNCTIONAL MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
q
CHEMISTRY HEATER ABNORMAL CONDITIONS AND PROTECTION . . . .15
q
TRAILING-EDGE TO TRAILING-EDGE FILM TIMES . . . . . . . . . . . . . . . . . . .15
q
HEATER PAD POWER CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
q
REPLENISHER RATES FOR DEVELOPER AND FIXER . . . . . . . . . . . . . . . .17
q
MEASURING REPLENISHER DISPENSING . . . . . . . . . . . . . . . . . . . . . . . . .18
PROCESSING STATE
SHUTTER STATE
MODE (Speed) CHANGE
or RESULT
Stand By
CLOSED
NORMAL
to
ENDO
to
NORMAL
Stand By
CLOSED
,
then
OPEN
for
4 seconds or more
Process starts, then stops at shutter closed and
stays in selected mode.
Stand By
CLOSED
,
then
OPEN
for
4 seconds or more
Process starts, and continues to operate in
selected mode.
Operating
OPEN
NORMAL
to
ENDO
to
NORMAL
Operating
CLOSED
ENDO
to
NORMAL
only
Operating ENDO
Cycle
CLOSED
,
then
OPEN
for
4 seconds or more
ENDO
will default to
NORMAL
then starts a
new cycle.
Operating Normal Cycle
CLOSED
,
then
OPEN
for
4 seconds or more
Remains in NORMAL mode,
and starts a new
cycle.
Cleaning (Manual)
CLOSED
or
OPEN
NORMAL
mode only.
Inhibits intial replenishment.
Shutter Action:
1. Accumunlates film length for replenishment.
2. Does not initiate process cycle.
End of ENDO Cycle
CLOSED
Restes to
NORMAL
mode
End of Normal Cycle
CLOSED
Stays in
NORMAL
mode
Power On
CLOSED
or
OPEN
Is in
NORMAL
mode
Power On
OPEN
Inhibits intial replenishment
FILM SHUTTER FUNCTIONS
Содержание A/T2000 XR
Страница 1: ...AUTOMATIC FILM PROCESSOR SERVICE MANUAL...
Страница 20: ...20 SYSTEM SCHEMATIC...
Страница 40: ...40 BASE ASSEMBLY TOP VIEW OVERHEAD VIEW DETAIL A BASE ASSEMBLY 45200...
Страница 42: ...42 BASE ASSEMBLY BOTTOM VIEW SINGLE FUSE CONFIGURATION 5A...
Страница 49: ...49 ROLLER TRANSPORTS WASH DRY TRANSPORT P N5510 DEVELOPER TRANSPORT P N45500 FIXER TRANSPORT P N 45505...