Page 42
VOLTAGE TRACKING
The power supply can be set so that the voltage of O/P2 is automatically set to that of O/P1 and
tracks any changes.
INTERFACES (MX180TP only)
Full digital remote control facilities are available through the USB, RS232, LAN and GPIB interfaces.
Setting and readback resolutions are the same as for the Output and Meter specifications
respectively.
RS232:
Standard 9-pin D-connector. Baud rate 9600.
USB:
Standard USB 2.0 hardware connection.
LAN:
Ethernet 100/10base-T hardware connection. 1.4 LXI Core 2011.
GPIB:
Conforms with IEEE488.1 and IEEE488.2.
Remote Command
Processing Time:
Typically 120ms (80ms min, 160ms max) between receiving the
command terminator for a step voltage change at the instrument and
the output voltage beginning to change.
GENERAL
AC Input:
115V – 240V AC ± 10%, 50/60Hz.
Installation Category II.
Power Consumption:
600VA max.
Operating Range:
+5ºC to +40ºC, 20% to 80% RH.
Storage Range:
−
40ºC to + 70ºC.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Safety & EMC:
Complies with EN61010-1 & EN61326-1.
For details, request the EU Declaration of Conformity for this
instrument via http://www.aimtti.com/support (serial no. needed).
Size:
210 x 130 x 375mm (WxHxD) half rack width x 3U height.
Weight:
5.8kg
Options:
19-inch rack kit for one or two units.
16
Default Values
When supplied from the factory the power supply is set as follows:
All Outputs
Voltage
1 volt
Current
0.1 amp
Range
30V/6A (O/P1 and O/P2); 5.5V/3A (O/P3)
OVP
140V (O/P1), 70V (O/P2), 14V (O/P3)
OCP
22A (O/P1), 12A (O/P2), 3.5A (O/P3)
Current meter averaging
Off (Level = Medium)
Voltage tracking
Off
Output On/Off
Off
Multi On/Off action
All Quick (synchronous on, synchronous off)
System Level Settings
Power-up state
Settings as at power-down, all outputs off
Beep sound
Enabled for key errors and trips
Disabled for entry into CC mode
Spin wheel action
Normal
The default values can be restored from the “Reset to factory Defaults” function – see section
9.7.