Page 40
15
Technical Specifications
General specifications apply for the temperature range 5°C to 40°C. Accuracy specifications apply
for the temperature range 18°C to 28°C after 15 minutes warm-up with no load and calibration at
23°C. Typical specifications are determined by design and are not guaranteed.
OUTPUT SPECIFICATIONS (Each Output)
Voltage/Current
Ranges:
O/P1: Switchable 0V to 30V/1mA to 6A, 0V to15V/1mA to 10A
or 0V to 60V/1mA to 3A.
Additional ranges of 0V to 30V/1mA to 12A, 0V to15V/1mA to 20A
or 0V to 60V/1mA to 6A, or 0V to 120V/1mA to 3A are available
with O/P2 disabled.
O/P2: 0V to 30V/1mA to 6A, 0V to15V/1mA to 10A
or 0V to 60V/1mA to 3A.
O/P3: 0V to 5.5V/10mA to 3A, 0V to 12V/10mA to 1.5A
Output Power:
More than to 375W total output power
Voltage Setting:
Resolution: O/P1 - 1mV (120V range 10mV); O/P2 - 1mV; O/P3 - 10mV
Accuracy: O/P1 - 0.05% ± 3mV, (120V range ± 6mV)
O/P2 - 0.05% ± 3mV
O/P3 - 0.3% ± 20mV
Current Setting:
Resolution: O/P1 and O/P2 – 1mA; O/P3 - 10mA
Accuracy: O/P1 - ± (0·3% + 3mA) to 3A, ± (0·5% + 3mA) to 10A
± (0·5% + 4mA) to 20A
O/P2 - ± (0·3% + 3mA) to 3A, ± (0·5% + 3mA) to 10A
O/P3 - ± (0·3% + 20mA)
Operating Mode:
Constant voltage or constant current with automatic cross-over.
CV or CC mode indication in display.
Output Switch:
Independent electronic switching with ON indication. In addition, Multi-On
and Multi-Off keys permit the outputs to be switched on/off synchronously.
Multi-On/Multi-Off
Action:
Individually settable delay between pressing of the Multi-On or Multi-Off key
and the turning on or off of the respective output. Delays settable between
10ms and 20 seconds. Separate delays for On and Off can be set.
Outputs can also be omitted from Multi-On or Multi-Off control.
Output Terminals:
Universal 4mm safety binding posts on 19mm (0·75”) spacing for Output;
screwless terminals for Sense.
MX180TP also has duplicate power and sense terminals at rear.
Ripple & Noise
CV Mode
(20MHz bandwidth):
For O/P 1, O/P 2 loaded at 15V/10A, O/P 3 loaded at 5.5V/3A, or for O/P 1
loaded at 15V/20A, 30V/12A, or 60V/6A (O/P 2 disabled):
All outputs typically <2mVrms, <15mV pk-pk; 3mVrms max.
O/P 1 loaded at 120V/3A: typically <3mVrms, <20mV pk-pk; 6mVrms max.
Load Regulation:
For any load change, measured at the output terminals, using remote sense:
Constant voltage: <0.01% ± 5mV
Constant current: <0.05% ± 1mA
Line Regulation:
Change in output for a 10% line change:
Constant voltage: <0.01% ± 5mV; Constant current: <0.01% ± 1mA
Transient Response:
<150us to within 50mV of set level for a 5% to 95% load change, except
O/P1 30V/12A, 15V/20A, 60V/6A and 120V/3A ranges <400us.
Temp. Coefficient:
Typically <100ppm/°C
Output Protection:
Outputs will withstand an applied forward voltage of up to:
O/P1 & O/P2 - 70V, O/P3 – 16V; O/P1 in 120V mode – 140V.
Reverse protection by diode clamp for reverse currents up to 3A, all outputs.
Over-voltage Protection
(OVP) Trip:
O/P1 & O/P2: 1V to 70V, typical response time 500us; O/P1 (120V mode):
1V-140V, typical response time 200ms; O/P3: 1V to 14V, typical response
time 500ms. Resolution 100mV, accuracy: ± (2% ±0.5V), all outputs.