9.6.Putting into service
To put the weather station into service the following steps must be taken. A connection overview is
shown in the Figure on page 21.
•
Cabling between ALMEMO
®
connector and the sensor
•
Feeding the necessary supply for heating and fan via external mains unit
(via mains unit plug or WAGO terminal strip)
•
Installation of overvoltage arrester (optional)
•
Completing the wiring and then plugging the ALMEMO
®
connector into
the measuring instrument
•
Checking the signal LEDs in the adapter
•
These LEDs and their respective meanings are listed in the following table.
LED
Description
POWER
Signals the status of the external supply voltage
LED active : External supply present
LED inactive : External supply absent or defective
ALMEMO
Signals the status of the ALMEMO
®
supply voltage from the instrument
LED active : ALMEMO
®
supply voltage active
LED inactive : ALMEMO
®
supply defective
FMD760
Signals the status of the sensor supply
LED active : Sensor supplied with 12 or 24 V
LED inactive : Sensor not supplied
Possible reasons : Defect in measuring instrument, mains unit, or sensor
Circuit-breaker in adapter is opened
9.7.Measuring quantities and ranges - factory default settings
Designation
Quantity
Expo-
nent
Measuring range Units
Resolution
1. Wind direction, averaged B-02
D Davg
0
0...+359 °
1 °
2. Wind velocity, averaged B-05
D Dvavg
0
0.5... 60.0 m/s
0.1 m/s
3. Wind velocity, maximum B-06
D Dvmax
0
0.5... 60.0 m/s
0.1 m/s
4. Atmospheric pressure
B-12
D p
0
600...1100.0 mbar
0.1 mbar
5. Temperature, at present B-09
D t
0
-52...+60.0 °C
0.1 K
6. Relative humidity
B-11
D Uw
0
0... 100.0 %rH
0.1 % rH
7. Rainfall quantity
B-13
D R
0
0...999.99 mm
0.01 mm
8. Rainfall intensity
B-15
D Ri
0
0...200.0 mm/h
0.1 mm/h
9.8.Configuration via the sensor menu
Measuring quantities and ranges configurable via ALMEMO
®
Control or the de-
vice itself
Combinations for the measuring channels can be configured from a list of 17 possible mea-
suring quantities and ranges. (* factory default settings):
- 22 -