Power supply
Heating
Fan
Rainfall RADAR
24/1 24 V via mains unit
Factory setting
Setting AUTO
Function Heating active
Setting AUTO
Function Fan active
Function
Continuous operation
24/2 24 V via mains unit
Setting OFF
Function
Heating deactivated
Setting AUTO
Function Fan active
Function
Continuous operation
24/3 24 V via mains unit
Setting OFF*
Function
Heating deactivated
(Energy saving mode 1)
Setting OFF*
Function
Fan deactivated
(Energy saving mode 1)
Function
Rainfall test every minute
If rainfall detected, conti-
nuously active
(Energy saving mode 1)
12/1 <12 V
via ALMEMO
®
device
130 mA
Power consumption
Setting AUTO
Function
Heating deactivated
(because no 24 V supply)
Setting AUTO
Function
Fan deactivated
(because supply <12 V)
Function
Continuous operation
12/2 <12 V
via ALMEMO
®
device
130 mA
Power consumption
Setting OFF
Function
Heating deactivated
Setting AUTO
Function
Fan deactivated
(because supply <12 V)
Function
Continuous operation
12/3 <12 V
via ALMEMO
®
device
25 mA
Power consumption
No rainfall test / no rain
130 mA
2 seconds per minute
for rainfall test
130 mA
Continuous if rain
Setting OFF*
Function
Heating deactivated
(Energy saving mode 1)
Setting OFF*
Function
Fan deactivated
(Energy saving mode 1)
Function
Rainfall test every minute
If rainfall detected, conti-
nuously active
(Energy saving mode 1)
*If fan setting is OFF, 'heating AUTO' is set automatically to 'heating OFF'.