13. Device configuration
Locking code VC
Functions
0
1
2
3
M , M
▲ ▼
Select measuring points, display measured value
●
●
●
●
M
▲
press and hold down Program double display
●
UBat display
●
●
●
●
10s
MAX-MIN-LIM select and display
●
●
●
●
10s
MAX-MIN delete
●
●
MEM save
●
●
●
MEM display and delete
●
●
30s
REL in the RAM
●
●
REL in the EEPROM as per sensor locking 4
≤
●
ADJ in the EEPROM as per sensor locking 5
≤
●
●
FCT Display sensor function as per sensor locking and ON
●
●
FCT Program sensor function as per sensor locking
●
FCT Display data logger function (with activated memory)
●
●
FCT Program data logger function (with activated memory)
●
13.3 Automatic device switch OFF
In menu item ´
AOFF
´ the device can be pro-
grammed to switch OFF automatically if no
key is touched for a certain settable number of
minutes; this will help save the batteries.
This automatic device switch OFF will not take
effect if the setting is ´- -´ or if a mains adapter
is connected.
13.4 Display illumination - automatic switch OFF
A characteristic of all illuminated displays is
that the illumination consumes a lot of power.
To prolong the device’s operating time it is ad-
visable therefore to switch off the illumination
either permanently or so long as the device is
not being used. The display on this device is a
static transflective color segment display; even
without illumination it can be clearly read but
the characters are only black. This operating mode is likewise selected when
the data logger is in sleep mode ´
SLPd
´. In normal operating mode, in menu
item ´
dOFF
´, the user can program the number of minutes of inactivity after
which display illumination is switched OFF automatically. This automatic
device switch OFF will not take effect if the setting is ´- -´ or if a mains adapter
is connected.