Data logger functions
Please note that alarm displays will increase current consumption
and, depending on the operating period, shorten the battery’s useful
life. To acknowledge - and thus terminate - alarm display when in
sleep mode press the
ON
key. The acoustic alarm signal will end
automatically after 10 seconds.
11.2.5 Date and time-of-day
For logging the measuring time a real-time
clock with date and time-of-day is incorpor-
ated. This real-time clock is buffered by the
power supply to the device; in the event of bat-
tery replacement date and time-of-day are
lost. Having selected the date and time-of-day
functions via key
FCT
information is entered
in the format specified in Section 9.3.
Function time-of-day ´
ZT
´ (
format:
hh:mm \ ss
)
Function date ´
DA
´ (
format:
dd.mon \ yyyy
)
After a battery change these func-
tions are automatically activated and
unlocked for the purpose of entering
information.
11.2.6 Measured value recording
Once the cycle, sleep mode (if applicable), and date and time-of-day have
been programmed, the measuring operation can be started in the measured
value display.
Before it starts the pause symbol is displayed in the status bar.
II
To start a cyclic measuring operation (cycle > 0) press
MEM
The start and save symbols then light up.
►
REC
In sleep mode ´
SLP
´ briefly appears and the device then switches off automat-
ically. Sleep mode can be terminated at any time by pressing
ON
The current measuring operation continues to run in normal mode.
The measuring operation can always be stopped by pressing
MEM
Individual manual measured value scans can be performed if cycle = 0.
In this case the symbol lights up briefly.
►
REC
11.2.7 Reading out the memory
The
internal measured value memory
can only be output via the serial inter-
face - preferably using the AMR-Control software.
(see Manual, 6.9.3.2)
With each output from the internal memory any one of the three output formats
ALMEMO
®
2470
35