Installation-, Operating-, Maintenance-Manual, Spare part List
MAK10
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Page 13
Translation - Manual MAK10 Rev. 11/2017
Components
Condensate-Discharge / Pump(s)
The sample gas is cooled down to +3°C in the heat exchanger. Water vapor condenses along the dew
point line as a result of the cooling. Condensate is separated from the sample gas and collected in the
condensate separator of heat exchanger. Condensate is continuously discharged by the connected
condensate pump (3,4ml/min @ 5U/min - N3,2x1,6mm).
FEATURES SR25
Self-suctioning peristaltic pump
Reliable continuous condensate removal
Synchronuous motor, gearbox with return stop
Slow speed, hose with long service life
Can operate in any installed position
Filter (optional)
In gas analytic systems occuring solid contaminants, especially fine particles up to 0,1 micron, are
reliably filtrated by this teflon depth filter. The large filter surface of ca.67 m² of the cylindrical
filtercartridge ensures a reliable micro-filtration and a long service life with low pressure drop. The
contamination level of the filter can be visually checked through the large porthole. A change of the
filtercartridge is extremely fast and easy to perform. The filter can operate in any installed position.
FEATURES TF70/TF90
Reliable filtration of solid particles (0,1 micron)
Optic check of the filtercartridge condition
Quick filtercartridge exchange
Small dead space
Can operate in any installed position