Installation-, Operating-, Maintenance-Manual, Spare part List
MAK10
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Page 21
Translation - Manual MAK10 Rev. 11/2017
Temperature-Monitoring and Error-Diagnosis
Operation Buttons
On/Off switch
Is used to switch the sample gas cooler on or off.
Temperature selector switch (TEMP)
Display of gas cooling temperature TLo or cooling air temperature TA.
Operation mode button (MODE)
For display and change of operation modes.
Displaying operation modes:
Press once = display of the operating hours during which a fault occurred. Last faults
can be displayed when button 2 is pressed one or more times.
Press twice = display of temperatures of the previously set fault. Different
temperatures of faults can be displayed when button 2 is pressed one or more times.
Press 3 times = mode for displaying the switch-on time of the condensate pumps.
Press 4 times = mode for displaying the switch-off time of the condensate pumps.
The TLo temperature is displayed again if a keypad input is not given within 10
seconds.
Changing operation modes:
Press the button once for 2 seconds = mode for changing the switch-on time of the
condensate pumps. The switch-on time of the condensate pump can be changed when
button 2 is pressed one or more times. Setting range 2 - 10 min.
Press the button once for 2 seconds and then press it once = mode for changing the
switch-off time of the condensate pumps. The switch-off time of the condensate pump
can be changed when button 2 is pressed one or more times. Setting range 0 - 10
min.
Press the button once for 2 seconds and then press it twice = mode for changing from
°C to °F. When button 2 is pressed once, the temperature display changes from °C to
°F or vice versa.
Press the button once for 2 seconds and then press trice = mode for changing the
switch-on and switch-off time of the condensate pumps to default settings.
Condensate pumps are reset to default settings by activating the mode for changing
the switch-on and switch-off time of the condensate pumps: 5 min On, 0 min Off.
The TLo temperature is displayed again if a keypad input is not given within 10
seconds.