![AGRI TRONIX Brecknell SBI-521 Series Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/agri-tronix/brecknell-sbi-521-series/brecknell-sbi-521-series_service-manual_2869497032.webp)
30
SBI521 Service Manual
7.
After
INP.PCT
is shown, 000000 (position of decimal point is determined by
CONFIG-FUNC-PERCEN setting) will be displayed
8.
Use the
[PRINT]
,
[N/G]
,
[UNIT]
keys to input the percentage of samples.
9.
Press the
[TARE]
key to confirm.
9a.
If the calculated unit-percentage-weight is less than 0.5d, the indicator will
display
Pct.Er
and go back to percent weighing mode. Otherwise, the scale
will go back to percent weighing mode. The unit percentage weight can be
saved after the power off and can be used next time.
3.6.3 Check Percent (percentage compare) in Percent Weighing Mode
To make percentage compare function be available, CONFIG-FUNC-COMPAR menu
item should set to YES, and high and low limitation of percentage should be set
correctly according to following steps:
1.
In percent weighing mode, Press the
[DATA]
key for more than 3 seconds to
enter compare data of high and low.
2.
After HIGH is shown,
000000
will be displayed. Use the
[PRINT]
,
[N/G]
,
[UNIT]
keys to input high percentage number.
3.
Press the
[TARE]
key to confirm. The HI annunciator will be illuminated.
4.
After Low is shown,
000000
will be displayed. use the
[PRINT]
,
[N/G]
,
[UNIT]
keys to input the low percentage number.
5.
Press the
[TARE]
key to confirm. The LO annunciator will be illuminated.
5a.
After a reasonable limitation is set and compare is active, one of annunciators
HI, OK, LO will be lit and the beeper will sound according to its setting in
USER-BEEP.
3.7 Hold Function
HOLD function can be used to freeze a displayed number. In this mode, the scale can
capture a dynamic number, hold a stable number, or average a unstable number and
then HOLD (freeze) this number temporarily for the user to view or record.
The HOLD function can be used in normal weighing mode, counting mode and percent
weighing mode. After entering HOLD mode, the speed of A/D converter can be
increased to 80Hz (if USER-HOLD-AD.H.SPD is set to YES) from original 10Hz for
some dynamic weighing applications.
It is possible to weigh restless weighing samples such as live animals or moving objects
within the hold function. The indicator provides special mode settings to accommodate
weight movements.
NOTE: If the High number value is 0 or is equal or less than low number, the
comparison will be disabled.