Vacuum Pump Isolation (VPI) Valve
21
To install:
1. Connect the 24 VDC power supply and the pump inverter to the mains power using the
splitter cord.
2. Connect the communication cable
DB9
connector into the inverter I/O port (
P1
).
3. Connect the 24 VDC power supply output cord to the communications cable labeled
24 V
.
4. Connect the VPI solenoid cord to the communications cable connector labeled
VPI
.