Technical Information
Window-Meanings
17
TPS-mini User Manual / 87-901-003-01 (E)
307/318
Window-Meanings
Tab. 8
N.
Read/
Write
Data
Type
Description
Admitted Values
000 R/W
L
Start/Stop
(in remote/ mode the
window is a read only)
Start = 1
Stop = 0
001 R/W
L
Low Speed Activation
(in remote/ mode the
window is a read only)
No = 0
Yes = 1
(default = 0)
008 R/W
L
Remote or Serial
configuration
Remote = 1
Serial = 0
(default = 1)
100 R/W
L
Soft Start
(write only in Stop condition)
YES = 1
NO = 0
101 R/W
N
R1 Set Point type
0 = Frequency
1 = Power
2 = Time
3 = Normal
4 =Pressure
(default = 0)
102 R/W
N
R1 Set Point valve
(expressed in Hz, W or s)
(default = 867)
103 R/W
N
Set Point delay: time
between the pump start and
the set point check
(seconds)
0 to 999999
(default = 0)
104 R/W
L
Set Point signal activation
type: the signal can be "high
level active" or "low level
active"
0 = high level active
1 = low level active
(default = 0)
105 R/W
N
Set point hysteresis
(in % of value)
0 to 100
(default = 2)