![Agilent Technologies TPS-mini 9699181 Скачать руководство пользователя страница 56](http://html1.mh-extra.com/html/agilent-technologies/tps-mini-9699181/tps-mini-9699181_user-manual_2868784056.webp)
Technical Information
Window Protocol
17
TPS-mini User Manual / 87-901-003-01 (E)
303/318
The controller can answers with the following response types:
Tab. 7
Type
Response
Length
Response
Value
Description
Logic 1
byte
-
After a read instruction of a logic
window
Numeric 6
bytes
-
After a read instruction of a numeric
window
Alphanumeric 10
bytes -
After a read instruction of an
alphanumeric window
ACK 1
byte
(0x6)
The command execution has been
successfully completed
NACK 1
byte
(0x15)
The command execution has been
failed
Unknown
Window
1 byte
(0x32)
The specified window in the
command is not a valid window
Data Type Error
1 byte
(0x33)
The data type specified in the
command (Logic, Numeric or
Alphanumeric) is not accorded with
the specified Window
Out of Range
1 byte
(0x34)
The value expressed during a write
command is out of range value of the
specified window
Win Disabled
1 byte
(0x35)
The specified window is Read Only or
temporarily disabled (for example) you
can’t write the Soft Start when the
Pump is running)