![Agilent Technologies TPS-mini 9699181 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/agilent-technologies/tps-mini-9699181/tps-mini-9699181_user-manual_2868784054.webp)
Technical Information
Window Protocol
17
TPS-mini User Manual / 87-901-003-01 (E)
301/318
Communication Protocol
The communication protocol is a MASTER/SLAVE type where:
Host = MASTER
Controller = SLAVE
The communication is performed in the following way:
1.
the host (MASTER) send a
M CRC
to the controller
(SLAVE);
2.
the controller answer with an
CRC
to the host.
The
MESSAGE
is a string with the following format:
<STX>+<ADDR>+<WIN>+<COM>+<DATA>+<ETX>+<CRC>
where:
NOTE
When a data is indicated between two quotes (‘...’) it means that the indicated
data is the corresponding ASCII character.
<STX>
(Start of transmission) = 0x02
<ADDR>
(Unit address) = 0x80 (for RS 232)
<ADDR>
(Unit address) = 0x80 + device number (0 to 31) (for RS
485)
<WIN>
(Window) = a string of 3 numeric character indicating the
window number (from ‘000’ to ‘999’); for the meaning of each
window see the relevant paragraph.
<COM>
(Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA>
= an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present. The field length is variable according to the data
type as per the following table
Содержание TPS-mini 9699181
Страница 3: ...TPS mini TPS mini User Manual 87 901 003 01 E 3 318 TPS mini...
Страница 19: ...16 Instructions for Use Disposal 266 318 TPS mini User Manual 87 901 003 01 E...
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 70: ......