Functions and Measurements
3
N9342C/43C/44C User’s Guide
55
Measuring Return Loss Using the Reflection
Calibration Routine
This procedure uses the reflection calibration
routine in the previous procedure
on page 53, to calculate the
return loss of the 370 MHz low- pass filter.
1
After calibrating the system with the above
procedure, reconnect the filter in place of the
short (cal device) without changing any analyzer
settings.
2
Use the marker to read return loss. Position the
marker with the front- panel knob to read the
return loss at that frequency.
Rotate the knob to find the highest peak and
the readout is the maximum return loss.
Figure 3-22 Measuring the Return Loss of the Filter