5
Key Reference
118
N9342C/43C/44C User’s Guide
• File Path
Allocate the path where the file is saved.
• File Size
Set the number of frames for the spectrogram trace
file. The allowed input range is 200 to 1500 frames.
Spectrum Emission Mask (SEM)
Spectrum Emission Mask measurement includes the
in- band and out- of- band spurious emissions. As it
applies to W- CDMA (3GPP), this is the power
contained in a specified frequency bandwidth at
certain offsets relative to the total carrier power. It
may also be expressed as a ratio of power spectral
densities between the carrier and the specified
offset frequency band. For WLAN, the reference
power is taken as the peak PSD in the signal, and
all offset results are also the peak PSD in that
offset as opposed to the integrated power.
Avg Number
Key access:
[MEAS] > {SEM} > {Avg Number}
Press
Avg Number (On)
to specify the number of
measurement averages used when calculating the
measurement result. The average will be displayed
at the end of each sweep.
Meas Type
Key access:
[MEAS] > {SEM} > {Meas Type}
Displays a menu where you can select a
measurement reference type,
Total Pwr Ref
or
PSD
Ref
. When
Total Pwr Ref
is selected, the power in the
carrier is used as the reference in computing the
relative power values for the offsets. When
PSD Ref
is selected, the mean power spectral density of the
carrier is used in computing the relative power
spectral density values for the offsets.
Содержание N9342C
Страница 1: ... Agilent N9342C 43C 44C Handheld Spectrum Analyzer User s Guide ...
Страница 8: ......
Страница 39: ...31 AgilentN9342C 43C 44C Handheld Spectrum Analyzer 3 Functions and Measurements ...
Страница 92: ...3 Functions and Measurements 84 N9342C 43C 44C User s Guide ...
Страница 156: ...5 Error Messages 148 N9342C 43C 44C User s Guide ...
Страница 178: ...7 Menu Map 170 N9342C 43C 44C User s Guide ...