Getting Started 3
Series N6700 User’s Guide
33
Sel
The command menu is now at the function control level. This is the lowest
level in this path.
Use the navigation keys to highlight the OVP Level control as shown below.
Enter the desired over-voltage level using the numeric keypad. Then press
Enter.
Channel
Press the Channel key at any time to select a different output channel. This
can save time because you can directly access the OVP control of each
channel without having to navigate through the menu levels.
NOTE
If you program an over-voltage protection level that is lower than the present
output voltage, the over-voltage protection circuit will trip and turn the output
channel off. The status indicator will show OV.
Exiting the Command Menu
There are two ways to exit the command menu.
Meter
Press the Meter key to immediately return to the metering screen. This is the
quickest way to return to metering mode.
Back
Press the Back key to back up one level at a time in the command menu.
This method may be more convenient if there are other menu commands to
be given.
Detailed instructions on how to use the power system’s functions and
capabilities are found in the next chapter. Detailed information about
the SCPI programming commands are found in chapters 5 and 6.
In Case of Trouble
Press the
Help
key to obtain additional help about any function
control menu level. Press the
Back
key to exit the Help menu.
The
Err
annunciator comes on if the self-test fails, or if other
operating problems occur with your instrument. Press the
Error
key
to display the list of errors. Refer to Appendix D for further
information.
The
SRQ
annunciator comes on if the instrument is requesting
service. Refer to Chapter 6 under “Status Subsystem” for more
information about the conditions that can generate a service request.