Installation Note N5230-90022 27
Reinstall the Outer Cover
CAUTION
This procedure is best performed with the analyzer resting on its front
handles in the vertical position.
Do not place the analyzer on its front panel
without the handles
. This will damage the front panel assemblies.
1. Slide the cover onto the frame.
2. Using a T-20 TORX driver, install the four rear panel feet (item
③
) by installing the center
screws (item
④
).
3. Slide the four bottom feet (item
⑤
) onto the cover.
4. Using a T-20 TORX driver, install the strap handles (item
①
) by tightening the screws
(item
②
) on both ends.
Figure 19 Outer Cover Reinstallation