Installation Note N5230-90022 25
Step 12. Reinstall the Front Panel Assembly
CAUTION
Before installing the front panel assembly onto the analyzer, place the analyzer
right side up and lift and support the front of the analyzer chassis.
Refer to
for this procedure.
1. Reconnect the ribbon cable (item
③
) to the A3 front panel interface board.
2. Slide the front panel over the test port connectors being careful to align the power switch
and floppy disk drive to their corresponding front panel cutouts. Ensure that the ribbon
cable (item
③
) is located below the fan to prevent it from being damaged by the fan blades.
3. With a T-10 TORX driver, install the eight screws (item
②
) in the sides of the frame.
4. Using a 5/16 inch torque wrench set to 10 in-lbs, install the six semi-rigid jumpers (item
①
) on the front panel.
Figure 17 Front Panel Assembly Reinstallation