Procedure for Declassifying a Faulty Instrument
Once the Processor and Hard Disk assemblies have been removed, proceed as in
Table 10-2
below:
Table 10-2
Assembly Disposal Procedures
Assembly
Procedure
Processor (CPU) Board
Either
Discard the processor board and send the instrument to a repair facility. A new
Processor Board will be installed, then the instrument will be repaired and
calibrated. If the instrument is still under warranty, you will not be charged for
the new Processor Board.
or
If you have another working instrument, install the Processor Board into that
instrument and erase the memory. Then reinstall the Processor Board back into
the non-working instrument and send it to a repair facility for repair and
calibration. If you discover that the Processor Board does not function in the
working instrument, discard the Processor Board and note that it caused the
instrument failure on the repair order. If the instrument is still under warranty,
you will not be charged for the new Processor Board.
A4 Memory Assembly
(MXG Vector
Instruments Only)
Either
Discard the board and send the instrument to a repair facility. A new board will
be installed, then the instrument will be repaired and calibrated. If the
instrument is still under warranty, you will not be charged for the new board.
or
If you have another working instrument, install the board into that instrument
and erase the memory. Then reinstall the board back into the non-working
instrument and send it to a repair facility for repair and calibration. If you
discover that the board does not function in the working instrument, discard the
board and note that it caused the instrument failure on the repair order. If the
instrument is still under warranty, you will not be charged for the new board.
Hard Disk
(E4438C Option 005,
E8267D Option 005, and
E8267C Option 005
only)
Either
Discard the Hard Disk and send the instrument to a repair facility. Indicate on
the repair order that the Hard Disk was removed and must be replaced. A new
Hard Disk will be installed, then the instrument will be repaired and calibrated.
If the instrument is still under warranty, you will not be charged for the new
hard disk.
or
Keep the Hard Disk and send the instrument to a repair facility. When the
instrument is returned, reinstall the Hard Disk.
Содержание N5171B
Страница 4: ...4 ...
Страница 6: ...6 Contents ...
Страница 8: ...8 Contacting Agilent Sales and Service Offices ...
Страница 12: ...12 Products Covered by this Document Recommended Software Upgrades ...
Страница 14: ...14 Security Terms and Definitions ...
Страница 38: ...38 Using Secure Display ...
Страница 42: ...42 Security Issues for Certain Firmware Revisions Recovering Erased System Files ...
Страница 46: ...46 Procedure for Declassifying a Faulty Instrument ...