29
Memory Clearing, Sanitization and Removal Procedures
6 Memory Clearing,
Sanitization and Removal
Procedures
This chapter describes several security functions you can use to remove sensitive data stored in the
instrument before moving it from a secure development environment. The functions described are:
•
“Erase All” on page 30
•
“Erase and Overwrite All” on page 31
•
“Erase and Sanitize All” on page 32
•
“Force Internal to SD Card” on page 33
•
“Clear Persistent State Information” on page 34
NOTE
The functions described in this chapter are
not
available in the E8257N instrument,
unless Option 340 has been installed.
CAUTION
These functions do
not
erase or sanitize external media connected to the instrument’s
USB port.
CAUTION
(This information applies only to
PSG
and
ESG
Signal Generators.)
Do not use the
Erase All
,
Erase and Overwrite All
,
Erase and Sanitize All
,
Erase
,
Overwrite
, or
Sanitize
procedure with any of these firmware revisions:
E44x8C: C.04.84, C.04.86, C.04.95
E82x7D: C.04.81, C.04.85, C.04.92
E8663B: C.04.90
If your instrument’s current firmware revision is among those listed above, please
upgrade
immediately
to newer firmware. For details, see
“Firmware Update
Procedure” on page 39
.
Содержание N5171B
Страница 4: ...4 ...
Страница 6: ...6 Contents ...
Страница 8: ...8 Contacting Agilent Sales and Service Offices ...
Страница 12: ...12 Products Covered by this Document Recommended Software Upgrades ...
Страница 14: ...14 Security Terms and Definitions ...
Страница 38: ...38 Using Secure Display ...
Страница 42: ...42 Security Issues for Certain Firmware Revisions Recovering Erased System Files ...
Страница 46: ...46 Procedure for Declassifying a Faulty Instrument ...