
40
Security Issues for Certain Firmware Revisions
Error Messages and Secure Environments
Error Messages and Secure Environments
If you cannot upgrade the firmware prior to using the security features, Agilent will help you to recover from
error messages that may appear after executing the security functions. The error messages indicate that
instrument system files have been erased. The following list shows some possible error messages:
•
256, File name not found; /SYS/LICENSE.TXT
•
617, Configuration Error; License file not found. Creating empty one.
•
A missing or damaged system file was encountered while trying to diagnose
the system.
Even if these error messages appear, the security function has completely sanitized the instrument. If the
instrument is located in a secure environment, it is safe to remove it. After removing it from the secure area,
follow the process below to recover the lost system files.
Содержание N5171B
Страница 4: ...4 ...
Страница 6: ...6 Contents ...
Страница 8: ...8 Contacting Agilent Sales and Service Offices ...
Страница 12: ...12 Products Covered by this Document Recommended Software Upgrades ...
Страница 14: ...14 Security Terms and Definitions ...
Страница 38: ...38 Using Secure Display ...
Страница 42: ...42 Security Issues for Certain Firmware Revisions Recovering Erased System Files ...
Страница 46: ...46 Procedure for Declassifying a Faulty Instrument ...