11
Products Covered by this Document
Recommended Software Upgrades
Recommended Software Upgrades
The information in this section applies only to
PSG
and
ESG
Signal Generators.
CAUTION
Do not use the
Erase All
,
Erase and Overwrite All
,
Erase and Sanitize All
,
Erase
,
Overwrite
, or
Sanitize
procedures with any of the firmware revisions listed in
Table 3-1
below.
If your instrument’s current firmware revision is among those listed, please upgrade
immediately
to newer firmware. For details, see
“Firmware Update Procedure” on
page 39
.
Table 3-1
Firmware revisions that require upgrades
Instrument model number
Current firmware revision
Upgrade to firmware revision
E44x8C
C.04.84, C.04.86, C.04.95
C.04.96 or later
E82x7D
C.04.81, C.04.85, C.04.92
C.04.94 or later
E8663B
C.04.90
C.04.93 or later