![Agilent Technologies MS40+ Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/agilent-technologies/ms40/ms40_user-manual_2868770081.webp)
Technical Information
RS 232/RS 485 Communication Description
5
MS40+ Single Stage Rotary Vane Pumps User Manual /
87-901-008-01 (
E)
81/90
Operational Limits
Tab. 10
INPUT VOLTAGE (V)
CONTROLLER STATUS
< 180
Power fail
180 – 264
Operative
> 264
Power fail
Status LED
Tab. 11
LED STATUS
CONTROLLER STATUS
Off
Stop
Green blinking
Ramp – Autotuning
Green
Normal Operation
Red
Fail
Red blinking
Oil level fail
Off
Reset + selftest
Orange blinking
Maintenance required
NOTE
When you switch the controller on an electronic self-test is performed, during
such phase you’ll see the led Orange for 1 sec. And turned off for two other
seconds. Immediately after the first phase the controller per-form a specific test
on the pump rotor in order to be sure that it isn’t locked.
Содержание MS40+
Страница 8: ...Contents 8 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 28: ...2 Gebrauchsanleitung Entsorgung 28 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 38: ...3 Mode d emploi Mise au rebut 38 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 48: ...4 Instructions for Use Disposal 48 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 84: ...5 Technical Information Accessories 84 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......