![Agilent Technologies MS40+ Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/agilent-technologies/ms40/ms40_user-manual_2868770067.webp)
Technical Information
Oil and Filter Cartridge Replacement Procedures
5
MS40+ Single Stage Rotary Vane Pumps User Manual /
87-901-008-01 (
E)
67/90
5.
Open the fill oil cap using a 10 mm hex key and add 1 liter of
SW60 oil.
6.
Close the oil fill cap tight.
7.
Replace the cartridge filter as per following slide.
Filter Cartridge Replacement Procedure
1.
Unscrew the cap rotating it anti-clock wise, by hand or socket
wrench with size 19 mm.
2.
Pull out the cartridge filter.
Figure 7
3.
Insert new cartridge filter and double check it is in the right
position (inserted in the appropriate hole).
4.
Tighten the cap rotating clock wise, by hand or socket wrench
with size 19 mm.
Cap
Cartridge Filter
P/N 9499201
Содержание MS40+
Страница 8: ...Contents 8 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 28: ...2 Gebrauchsanleitung Entsorgung 28 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 38: ...3 Mode d emploi Mise au rebut 38 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 48: ...4 Instructions for Use Disposal 48 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 84: ...5 Technical Information Accessories 84 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......