Technical Information
RS 232/RS 485 Communication Description
5
MS40+ Single Stage Rotary Vane Pumps User Manual /
87-901-008-01 (
E)
75/90
RS 232/RS 485 Communication Description
Both the RS 232 and the RS 485 interfaces are available on the
connector J2
J2 – Serial Port
Tab. 6
PIN N.
SIGNAL NAME
1
+ 5 V out
2
TX (RS232)
3
RX (RS232)
4
Spare
5
GND
6
A+ (RS485)
7
Spare
8
B- (RS485)
9
Spare
The communication protocol is the same (see the structure below),
but only the RS 485 manages the address field. Therefore to enable
the RS 485 is necessary to select the type of communication as well
as the device address by means of the T-plus software.
NOTE
The RS-485 is a 2-wire (gnd optional) half-duplex communication link.
Communication Format
8 data bit
no parity
1 stop bit
baud rate: 600/1200/2400/4800/9600 programmable
Содержание MS40+
Страница 8: ...Contents 8 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 28: ...2 Gebrauchsanleitung Entsorgung 28 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 38: ...3 Mode d emploi Mise au rebut 38 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 48: ...4 Instructions for Use Disposal 48 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 84: ...5 Technical Information Accessories 84 90 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 E...
Страница 85: ......
Страница 86: ......
Страница 87: ......
Страница 88: ......
Страница 89: ......