Troubleshooting
4
7000 Series Oscilloscopes Service Guide
73
6
If you still do not see the waveform, use the
troubleshooting flowchart in this chapter to isolate the
problem.
To compensate the analog probes
You should compensate your analog probes to match their
characteristics to the oscilloscope’s channels. A poorly
compensated probe can introduce measurement errors.
1
Perform the procedure
“To verify basic oscilloscope
operation”
on page 72
2
Use a nonmetallic tool to adjust the trimmer capacitor on
the probe for the flattest pulse possible. The trimmer
capacitor is located on the probe BNC connector.
Figure 11
Example pulses
3
Connect probes to all other analog channels (channel 2 of
a 2- channel oscilloscope, or channels 2, 3, and 4 of a
4- channel oscilloscope). Repeat the procedure for each
channel. This matches each probe to each channel.
The process of compensating the probes serves as a basic
test to verify that the oscilloscope is functional.
comp.cdr
Perfectly compensated
Over compensated
Under compensated