Testing Performance
2
7000 Series Oscilloscopes Service Guide
35
c
Press the
Acquire
key. Then press the
Averaging
softkey
and set
#Avgs
to 64.
Wait a few seconds for the measurement to settle.
2
Press the
Cursors
key, set the
Mode
softkey to
Normal
, then
press the
X Y
softkey and select
Y
. Press the
Y1
softkey,
then use the Entry knob (labeled
on the front panel)
to set the Y1 cursor on the baseline of the signal.
3
Use the BNC tee and cables to connect the oscilloscope
calibrator /power supply to both the oscilloscope and the
multimeter (see
Figure 3
).
Table 7
Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting
Oscilloscope
Calibrator Setting
Test Limits
5 V/Div
35 V
34.04 V
to
35.96 V
2 V/Div
14 V
13.616 V
to
14.384 V
1 V/Div
7 V
6.808 V
to
7.192 V
500 mV/Div
3.5 V
3.404 V
to
3.596 V
200 mV/Div
1.4 V
1.3616 V
to
1.4384 V
100 mV/Div
700 mV
680.8 mV
to
719.2 mV
50 mV/Div
350 mV
340.4 mV
to
359.6 mV
20 mV/Div
140 mV
136.16 mV
to
143.84 mV
10 mV/Div
70 mV
68.08 mV
to
71.92 mV
5 mV/Div
35 mV
34.04 mV
to
35.96 mV
2 mV/Div
1
14 mV
13.232 mV
to
14.768 mV
1
Full scale is defined as 32 mV on the 2 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
Содержание InfiniiVision 7000 Series
Страница 1: ...s1 Agilent InfiniiVision 7000 Series Oscilloscopes Service Guide ...
Страница 8: ...8 Service Guide Contents ...
Страница 12: ...12 Service Guide Tables ...
Страница 20: ...20 7000 Series Oscilloscopes Service Guide 1 Characteristics and Specifications ...
Страница 109: ...Replaceable Parts 6 7000 Series Oscilloscopes Service Guide 109 Figure 42 Exploded View 2 of 2 ...