72
7000 Series Oscilloscopes Service Guide
4
Troubleshooting
To verify basic oscilloscope operation
1
Press the
Save/Recall
key on the front panel, then press
the
Default Setup
softkey under the display. The oscilloscope
is now configured to its default settings.
2
Connect an oscilloscope probe from channel 1 to the
Probe
Comp
signal terminal on the front panel.
3
Connect the probe’s ground lead to the ground terminal
that is next to the
Probe Comp
terminal.
4
Press
AutoScale
.
5
You should see a waveform on the oscilloscope’s display
similar to this:
If you see the waveform, but the square wave is not
shaped correctly as shown above, perform the procedure
“To compensate the analog probes”
on page 73.
If you do not see the waveform, ensure your power source
is adequate, the oscilloscope is properly powered- on, and
the probe is connected securely to the front- panel analog
channel input BNC and to the Probe Comp terminal.