Testing Performance
2
7000 Series Oscilloscopes Service Guide
39
1
Connect the equipment (see
Figure 5
).
a
Use the N cable to connect the signal generator to the
input of the power splitter input.
b
Connect the power sensor to one output of the power
splitter.
c
Use an N- to- BNC adapter to connect the other splitter
output to the channel 1 input.
Table 8
Equipment Required to Verify Bandwidth
Equipment
Critical Specifications
Recommended
Model/Part
Signal Generator
100 kHz - 1 GHz at 200 mVrms Agilent N5181A,
E4400B, or 8648A
Power Meter/Sensor 1 MHz - 1 GHz ±3% accuracy
Agilent E4418B/8482A
Power Splitter
outputs differ by < 0.15 dB
Agilent 11667A
Cable
Type N (m) 24 inch
Agilent 11500B
Adapter
Type N (m) to BNC (m)
Agilent 1250-0082 or
Pomona 3288 with
Pomona 3533