48
7000 Series Oscilloscopes Service Guide
2
Testing Performance
Test limits for:
•
Internal trigger on all models:
•
< 10 mV/div: greater of 1 div or 5 mV
pp
>=10 mV/div: 0.6 div
•
External trigger on all 2- channel models
(MSO/DSO7xx2A/B):
•
Trigger range: 1.0V
DC to 100 MHz: < 100 mV
pp
>100 MHz to max bandwidth: < 200 mV
pp
•
Trigger range: 8.0V
DC to 100 MHz: < 250 mV
pp
>100 MHz to max bandwidth: < 500 mV
pp
•
External trigger on 4- channel models:
(MSO/DSO7xx4A/B)
DC to 500 MHz: < 500 mV
pp
Table 11
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal
Generator
25 MHz, 100 MHz, 350 MHz,
500 MHz, and 1 GHz sine
waves
Agilent N5181A, E4400B, or
8648A
Power splitter Outputs differ < 0.15 dB
Agilent 11667A
Cable
BNC, Qty 3
Agilent 10503A
Adapter
N (m) to BNC (f), Qty 3
Agilent 1250-0780
Feedthrough
50
Ω
BNC (f) to BNC (m)
Agilent 0960-0301