16-13
Function Commands
FFT:RESolution?
FFT:RESolution?
Query
:FUNCtion<N>:FFT:RESolution?
The :FUNCtion<N>:FFT:RESolution? query returns the current resolution of
the FFT function.
Returned Format
[FUNCtion<N>:FFT:RESolution] <resolution_value><NL>
<N>
An integer from 1 to 4 representing the selected function.
<resolution
_value>
Resolution frequency.
The FFT resolution is determined by the sample rate and memory depth
settings. The FFT resolution is calculated using the following equation:
FFT Resolution = Sample Rate / Effective Memory Depth
The effective memory depth is the highest power of 2 less than or equal to the
number of sample points across the display. The memory bar in the status area
at the top of the display indicates how much of the actual memory depth is
across the display.
Содержание Infiniium 8000A
Страница 1: ...Agilent Technologies Infiniium 8000A Programmer s Reference ...
Страница 2: ......
Страница 20: ...Contents 16 ...
Страница 21: ...1 Introduction to Programming ...
Страница 43: ...2 LAN and GPIB Interfaces ...
Страница 53: ...3 Message Communication and System Functions ...
Страница 58: ...3 6 ...
Страница 59: ...4 Status Reporting ...
Страница 64: ...4 6 Status Reporting Status Reporting Data Structures Figure 4 2 Status Reporting Data Structures ...
Страница 78: ...4 20 Figure 4 3 Status Reporting Decision Chart ...
Страница 79: ...5 Programming Conventions ...
Страница 84: ...5 6 Programming Conventions The Command Tree Figure 5 1 Command Tree ...
Страница 85: ...5 7 Programming Conventions The Command Tree Figure 5 2 Command Tree Continued ...
Страница 86: ...5 8 Programming Conventions The Command Tree Figure 5 3 Command Tree Continued ...
Страница 87: ...5 9 Programming Conventions The Command Tree Figure 5 4 Command Tree Continued ...
Страница 88: ...5 10 Programming Conventions The Command Tree Figure 5 5 Command Tree Continued ...
Страница 89: ...5 11 Programming Conventions The Command Tree Figure 5 6 Command Tree Continued ...
Страница 94: ...5 16 ...
Страница 95: ...6 Sample Programs ...
Страница 149: ...7 Acquire Commands ...
Страница 176: ...7 28 Acquire Commands SRATe AUTO ...
Страница 177: ...8 Bus Commands ...
Страница 187: ...9 Calibration Commands ...
Страница 195: ...10 Channel Commands ...
Страница 223: ...11 Common Commands ...
Страница 247: ...12 Digital Commands ...
Страница 254: ...12 8 ...
Страница 255: ...13 Disk Commands ...
Страница 300: ...13 46 Disk Commands STORe Obsolete ...
Страница 301: ...14 Display Commands ...
Страница 322: ...14 22 ...
Страница 323: ...15 External Trigger Commands ...
Страница 343: ...16 Function Commands ...
Страница 382: ...16 40 ...
Страница 383: ...17 Hardcopy Commands ...
Страница 391: ...18 Histogram Commands ...
Страница 403: ...19 InfiniiScan ISCan Commands ...
Страница 421: ...20 Limit Test Commands ...
Страница 429: ...21 Marker Commands ...
Страница 452: ...21 24 ...
Страница 453: ...22 Mask Test Commands ...
Страница 499: ...23 Measure Commands ...
Страница 636: ...23 138 Measure Commands VUPPer ...
Страница 637: ...24 Pod Commands ...
Страница 642: ...24 6 ...
Страница 643: ...25 Root Level Commands ...
Страница 645: ...25 3 STORe SETup STORe WAVeform TER Trigger Event Register VIEW ...
Страница 674: ...25 32 ...
Страница 675: ...26 Self Test Commands ...
Страница 679: ...27 System Commands ...
Страница 694: ...27 16 ...
Страница 695: ...28 Time Base Commands ...
Страница 708: ...28 14 ...
Страница 709: ...29 Trigger Commands ...
Страница 822: ...29 114 ...
Страница 823: ...30 Waveform Commands ...
Страница 893: ...31 Waveform Memory Commands ...
Страница 902: ...31 10 ...
Страница 903: ...32 Error Messages ...
Страница 914: ...32 12 ...