![Agilent Technologies G6011A Quiet Cover MS Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/agilent-technologies/g6011a-quiet-cover-ms/g6011a-quiet-cover-ms_service-manual_2867747014.webp)
14
Quiet Cover MS Service Manual
2
Replacement Procedures
Preparing the Quiet Cover MS for Repair
Before performing any replacement procedures on the Quiet
Cover MS, prepare it for service as described below.
Required tools
•
5/16- inch open- end wrench
Procedure
1
Turn off and unplug the Quiet Cover MS.
2
Remove the two pump access panels by loosening the
thumbscrews on the top and sides.
3
Remove the lid by loosening the four thumbscrews on
each side.
4
Remove the pump.
a
Disconnect the power cord from the rough pump (see
).
WA R N I N G
The pump:
• May be hot. Use gloves when moving it or wait until it is cool.
• Is heavy. For your safety, a two-person lift is required. Failure
to perform a two-person lift can result in injury.
Figure 4
Pump power cord (G6011A Series)
Pump power cord