Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
10
FID
– Remove four mounting screws.
TCD
– Remove two mounting screws.
6.
Lift the detector and flow manifold out of the lid as a unit.
WARNING
The fibrous insulation material can cause irritation to the skin, eyes, and
mucous membranes. Always wear gloves when working with the insulation.
If the insulation is flaky/crumbly, wear protective eyewear and respirator.
7.
Remove any insulation that may be in the hole under the detector.
Mounting screws
(one concealed)
Mounting
screws