Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
8
TCD
– Unplug the two-wire switching valve cable at the connector. To
release the PRT and filament wires, press down on the orange tabs. Pass
the disconnected wires through the hole to the top side of the lid.
7.
Lower the stop plate on the lid cam. Pull the lid forward until the cam
follower rests on the curved surface of the cam. Leave the stop plate in
the "down" position.
8.
Close the lid.
Ribbon cable connector
Ignitor cable connector
Switching valve cable connector
PRT wires
Filament wires