Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
9
Remove the Existing Detector and Flow Manifold
1.
On the top of the lid, loosen the connector cover plate screws and
remove the plates.
2.
Disconnect the gas fittings. The figure shows the FID manifold; the TCD
manifold is similar but has only one gas connection.
3.
Trace the ribbon cable from the flow manifold and disconnect it from
the harness.
4.
Remove the three manifold mounting screws.
5.
Remove the screws holding the detector assembly in place.
Cover plate screws (4)
Manifold mounting
Gas fittings
Gas manifold
(FID shown)
screws