Installing a Flame Ionization Detector on an Agilent 6850 Gas Chromatograph
11
Install the Flame Ionization Detector and Flow Manifold
1.
Place the flame ionization detector and flow manifold in position in the
lid. Check that no wiring or tubing is trapped under either assembly.
2.
Secure the detector assembly to the lid with four screws.
3.
Secure the manifold to the lid with three screws.
4.
Plug the manifold ribbon cable into its connector.
5.
Install the two connector cover plates.
6.
Connect the gas supply lines as shown below. If the previous detector
was a TCD, the
A
and
H
lines should be folded up in the space under the
fittings. Unfold them, being careful to avoid kinks, and feed them up to
the manifold area.
Screws (4 places,
Ignitor cable
Heater/sensor cable connection
one not visible)
Connect the line marked H here
Connect the line marked M here
Connect the line marked A here