![Agilent Technologies G2545A Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/agilent-technologies/g2545a/g2545a_calibration-manual_2867736007.webp)
Step 3. Retrieve the temperature data from the digital recording thermometer
Hybridization Oven Calibration Procedure
7
Step 3. Retrieve the temperature data from the
digital
recording thermometer
•
Follow the instructions for the
digital recording thermometer
to retrieve
the temperature data.
These steps are for the
Fourtec MicroLite II USB
digital recording
thermometer
.
1
Start the
Fourtec DataSuite Software
.
2
Remove the cover on the
Fourtec MicroLite II USB
digital recording
thermometer
and insert it into a USB port on the computer.
3
Right-click the
Logger
icon and click
Stop
to stop data recording.
4
Right-click the
Logger
icon and click
Download Data.
The
Fourtec DataSuite Software
automatically saves the data on the
computer.
5
Double-click the
Logger
icon to view a graph of the data.
6
To view archived data:
a
Click
File > Open
.
b
In the Open Data Files dialog box, select the time period and logger
for which you want to view the data.
c
Click
OK
.
The archived data appears in History View. The Data File Path shows
the folder in which the logger data was saved.
7
To export the data to a spreadsheet, click
Export to Excel
in the
lower graph toolbar.
8
Record the oven temperature from the
digital recording thermometer
.