Agilent G2545A Hybridization Oven
Calibration Procedure
Step 1. Set up and heat the oven
3
Step 2. Start the digital recording thermometer
4
Step 3. Retrieve the temperature data from the digital recording
thermometer
7
Step 4. Calibrate the oven
9
The Agilent G2545A Hybridization Oven maintains temperature calibration
for at least 3 months under normal operating conditions when properly
installed within the laboratory.
See the
Agilent G2545A Hybridization Oven Installation, Operation, and
Maintenance Guide
(p/n G2545-9001) for information on site
requirements and oven placement.
Calibrate the Agilent G2545A Hybridization Oven:
•
After first installation in a working environment
•
After each 3 months of use
•
When systematic high microarray background noise may indicate a drift
in oven temperature
Do the calibration steps in this guide to maintain optimal performance for
your Agilent G2545A Hybridization Oven.
To view a demonstration of the hybridization oven
calibration procedure, go to
http://www.agilent.com/genomics/protocolvideos