![Agilent Technologies G2545A Скачать руководство пользователя страница 2](http://html.mh-extra.com/html/agilent-technologies/g2545a/g2545a_calibration-manual_2867736002.webp)
2
Hybridization Oven Calibration Procedure
Required Parts and Tools
Part
Vendor and part/model
hybridization oven
Agilent p/n G2545A
hybridization oven rotator rack
Agilent p/n G2530-60029
hybridization chambers
Agilent p/n G2534A
Fourtec MicroLite II USB digital recording
thermometer or equivalent
•
Range -40°C to 80°C
•
Resolution 0.1°C
•
Accuracy ±0.3°C
MicroDAQ.com
•
p/n LITE5008P (8,000 readings)
•
p/n LITE5032P (32,000 readings)
Computer and software to work with the
digital
recording thermometer
Fourtec DataSuite Software or other software
as specified by the
digital recording
thermometer
vendor
Calibration service to 65°C for
digital recording
thermometer
MicroDAQ.com p/n CAL-NIST-TEMP or other
service as recommended by
digital recording
thermometer
vendor
general-purpose labeling tape
or
2 cable ties (long enough to attach the digital
recording thermometer to the oven rotator rack)