Step 2. Start the digital recording thermometer
Hybridization Oven Calibration Procedure
5
6
Set
Temperature Scale
to
°C
.
7
Mark the
Timer Run
check box
and select the date and time to
start the recorder.
8
Set
Interval
to
00:05:00
(5
minutes).
9
Click
Setup
.
10
Click
Yes
to continue.
11
Remove the unit from the USB
port and replace the cover.
12
Put the
digital recording
thermometer
onto the middle of
the
hybridization oven rotator
rack
.
13
Secure the
digital recording
thermometer
to the
hybridization
oven rotator rack
. Use
general-purpose laboratory
labeling tape or cable ties.