Chapter 2
77
Programming Fundamentals
SCPI Language Basics
of commands indicated below:
[:SENSe]
:POWer
[:RF]
:ATTenuation 40dB
:TRIGger
[:SEQuence]
:EXTernal [1]
:SLOPe
POSitive
[:SENSe]
:FREQuency
:STARt
:POWer
[:RF]
:MIXer
:RANGe
[:UPPer]
Bad Command
Good Command
PWR:ATT 40dB
POW:ATT 40dB
The short form of
POWER
is
POW
, not
PWR
.
FREQ:STAR 30MHz;MIX:RANG -20dBm
FREQ:STAR 30MHz;POW:MIX:RANG -20dBm
The
MIX:RANG
command is in the same
:SENSE
subsystem as
FREQ
, but executing the
FREQ
command puts you back at the
SENSE
level. You must specify
POW
to get to the
MIX:RANG
command.
FREQ:STAR 30MHz;POW:MIX RANG -20dBm
FREQ:STAR 30MHz;POW:MIX:RANG -20dBm
MIX
and
RANG
require a colon to separate them.
:POW:ATT 40dB;TRIG:FREQ:STAR 2.3GHz
:POW:ATT 40dB;:FREQ:STAR 2.3GHz
:FREQ:STAR
is in the
:SENSE
subsystem, not the
:TRIGGER
subsystem.
:POW:ATT?:FREQ:STAR?
:POW:ATT?;:FREQ:STAR?
:POW
and
FREQ
are within the same
:SENSE
subsystem, but they are two separate commands, so
they should be separated with a semicolon, not a colon.
:POW:ATT -5dB;:FREQ:STAR 10MHz
:POW:ATT 5dB;:FREQ:STAR 10MHz
Attenuation cannot be a negative value.