Chapter 1
47
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Negative Transition
This command determines which bits in the Questionable
Condition register will set the corresponding bit in the Questionable Event register when
the condition register bit has a negative transition (1 to 0). The variable <integer> is the
sum of the decimal values of the bits that you want to enable.
Questionable Positive Transition
This command determines which bits in the Questionable
Condition register will set the corresponding bit in the Questionable Event register when
the condition register bit has a positive transition (0 to 1). The variable <integer> is the
sum of the decimal values of the bits that you want to enable.
Questionable Calibration Register
Questionable Calibration Condition
This query returns the decimal value of the sum of the
bits in the Questionable Calibration Condition register.
The data in this register is continuously updated and reflects the current conditions.
Mode
All
Remote Command
:STATus:QUEStionable:NTRansition 16 Temperature
summary ‘questionable cleared’ will be reported to the Status
Byte Register.
:STATus:QUEStionable:NTRansition?
Example
STAT:QUES:NTR 16 Temperature summary ‘questionable
cleared’ is reported to the Status Byte Register.
Preset
0
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:PTRansition <integer>
:STATus:QUEStionable:PTRansition?
Example
STAT:QUES:PTR 16 Temperature summary ‘questionable
asserted’ will be reported to the Status Byte Register.
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All