
Chapter 1
67
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Temperature Register
Questionable Temperature Condition
This query returns the decimal value of the sum of the
bits in the Questionable Temperature Condition register.
NOTE
The data in this register is continuously updated and reflects the
current conditions.
Questionable Temperature Enable
This command determines which bits in the Questionable
Temperature Condition Register will set bits in the Questionable Temperature Event
register, which also sets the Temperature Summary bit (bit 4) in the Questionable
Register. The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Questionable Temperature Event Query
This query returns the decimal value of the sum of
the bits in the Questionable Temperature Event register.
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:CONDition?
Example
STAT:QUES:TEMP:COND?
Preset
0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode
All
Remote Command
:STATus:QUEStionable:TEMPerature:ENABle <integer>
:STATus:QUEStionable:TEMPerature:ENABle?
Example
STAT:QUES:TEMP:ENAB 1 Reference Oscillator Oven Cold
will be reported to the Temperature Summary of the Status
Questionable register.
Preset
32767
Min
0
Max
32767
SCPI Status Bits/OPC
Dependencies
Sequential command