Measurement Synchronization with LXI-B
Setting Up and Executing a List Sweep Measurement
Synchronizing Using Alarm Triggers
A second method of synchronization uses alarm triggers to step each
instrument through their lists at predetermined times. Using this
method, the instruments never have to communicate their states.
Instead, the controller determines a schedule for execution, and each
instrument carries out its actions with the assumption that the other is
keeping up with its schedule. The following example illustrates a
method for setting up a list sweep using alarms for synchronization.
31. Begin the MXA’s list
sweep
:INIT:LIST
This command is sent
to the MXA.
This will cause the
MXA to take a
measurement at the
first frequency in its
list. It will also begin
waiting for the alarm to
fire before moving on to
the next measurement.
32. Wait for the
“OperationComplete”
instrument event
Detect the peer to peer traffic
using Interactive LXI. To
programmatically listen for LXI
peer to peer messages, use the
IOLibraries TMFramework LXI
library.
When the MXA has
completed its list, it will
send the peer to peer
message
“OperationComplete”
with a rising edge.
33. Read measurement
results.
:FETCH:LIST?
This query is sent to
the MXA.
Task
Action
Notes
1. Disable LXI Output
LAN Events on the
MXG
:LXI:EVENt:OUTPut:LAN:DISable:ALL
This command is sent
to the MXG
2. Reset the MXG
*RST
This command is sent
to the MXG
3. Choose the MXG’s
PTP domain
:LXI:CLOCk:PTP:DOMain 0
This command is sent
to the MXG.
The parameter may