98
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
Informational
<filename> file loaded
The filename indicated has been successfully loaded.
Informational
<filename> file saved
The filename indicated has been successfully saved.
Informational
<filename> too many data entries
This message may appear when loading data from a limit line or
amplitude correction file. The [DATA] section of such a file can contain at
most 200 lines of data. This message is displayed if that limit is exceeded.
Informational
<filename1> file renamed to
<filename2>
<name1> file renamed to <name2>
File name1 has been successfully renamed to file name2.
Informational
Atten auto set to 15 dB
Indicates that an input signal has been detected which is of sufficient level
to damage the input circuitry and the input attenuator has been
automatically set to 15 dB. If the signal level is reduced, the attenuator
will stay at 15 dB. This overload protection occurs at an input power level
of 13 dBm (68 dBmV for Option 1DP)
±
7 dB when the input attenuation is
auto coupled and set to <15 dB. To return to the original measurement
setup, reduce the input signal level and press
Amplitude
. Then press
Attenuation (Auto)
.
Overload protection is only available in the Agilent Technologies E4401B
and E4411B.
Informational
Auto ranging...
Displayed during autoranging.
Informational
B7D and/or B7E not found. Code Domain not available.
Digital Signal Processing and Fast Analog to Digital Converter (Option
B7D) and/or RF Communications Hardware (Option B7E) are not installed
options on your analyzer. Code domain is therefore not available.
Informational
B7D and/or B7E not found. Mod Acc not available.
Digital Signal Processing and Fast Analog to Digital Converter (B7D)
and/or RF Communications Hardware (B7E) are not installed options on
your analyzer. Modulation accuracy is therefore not available.
Informational
Carrier Not Present.
A carrier signal/burst is expected at the analyzer input. This signal cannot
be found; however, the measurement will still run. Check input signal
connection.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...