102
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
Informational
WARNING: You are about to delete all of the contents of
directory “x:\xxxxx\”. Press Delete Now again to proceed or
any other key to abort.
WARNING: You are about to delete all of the contents of
directory xxxxxx. Press Delete Now again to proceed or any
other key to abort.
If you select a directory or subdirectory to delete, this popup message is
displayed when you press
Delete Now
. (xxxxx in the message is the full
path and directory name).
Informational
Waiting for end of data acquisition to access floppy disk
Informational
Volume <name> formatted
The indicated disk has been successfully formatted.
Informational
Time resolution is low, reduce Meas Time or increase Trace
Points
Changing either Meas Time or Trace Points has increased the burst to
mask uncertainty greater than 0.4 symbol. You should lower the Meas
Time or increase Trace Points to reduce this uncertainty.
Status Messages
These messages indicate conditions that can cause the display of incorrect data. The name
of the corresponding status bit appears in parenthesis. Some messages display
only
the
status bit (as noted).
Status messages appear in the upper right portion of the screen grid.
Status
* (Invalid Data)
The * (Invalid Data) is displayed in the upper right-hand corner of the
graticule to let you know that the display does not match the current
settings. When doing a
Max Hold
, then switching to
View
, the asterisk will
appear. This is due to the analyzer’s auto-coupling rules (the Video
Bandwidth and detector type will change when
View
is pressed). To avoid
this situation, change the
Video BW
to
Man
(manual) and change the
detector type to
Peak
before pressing
Max Hold
.
Status
50 MHz Osc Unlevel (50 MHz Osc Unleveled)
The internal 50 MHz amplitude reference source has become unleveled.
This condition must be corrected before a valid alignment can be
performed.
Status
ADC Align Failure
status bit only, no message
The alignment routine was unable to align the analog-to-digital converter
(ADC).
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...