Frequency Readout Accuracy
ESA
Functional T
ests
Figure 4-5 Equipment Setup
Figure 4-6 Equipment Setup
Procedure
1. Connect the equipment as shown in
or
. Remember to connect the
10 MHz REF OUT of the analyzer to the 10 MHz REF IN of the synthesized sweeper.
2. Perform the following steps to set up the equipment:
a. Press
INSTRUMENT PRESET
on the synthesized sweeper, then set the controls as
follows:
CW
, 1.490,
GHz
POWER LEVEL
, 10,
– dBm